An optical time-domain reflectometer (OTDR) is an optoelectronic instrument used to characterize an optical fiber. An OTDR is the optical equivalent of an electronic time domain reflectometer. It injects a series of optical pulses into the fiber under test. It also extracts, from the same end of the fiber, light that is scattered (Rayleigh backscatter) or reflected back from points along the fiber. (This is equivalent to the way that an electronic time-domain reflectometer measures reflections caused by changes in the impedance of the cable under test.) The strength of the return pulses is measured and integrated as a function of time, and plotted as a function of fiber length.

Description[edit]

OTDR is short for Optical Time-Domain Reflectormeter

An OTDR may be used for estimating the fiber length and overall attenuation, including splice and mated-connector losses. It may also be used to locate faults, such as breaks, and to measure optical return loss. To measure the attenuation of multiple fibers, it is advisable to test from each end and then average the results, however this considerable extra work is contrary to the common claim that testing can be performed from only one end of the fiber.

In addition to required specialized optics and electronics, OTDRs have significant computing ability and a graphical display, so they may provide significant test automation. However, proper instrument operation and interpretation of an OTDR trace requires special technical training and experience.

OTDRs are commonly used to characterize the loss and length of fibers as they go from initial manufacture, through to cabling, warehousing while wound on a drum, installation and then splicing. The last application of installation testing is more challenging, since this may involve extremely long distances, or multiple splices spaced at short distances, or fibers with different optical characteristics joined together. OTDR test results are often stored in case of later fiber failure or warranty claims. Fiber failures can be very expensive, both in terms of repair cost, and service loss.

OTDRs are also commonly used for fault location on installed systems. In this case, reference to the OTDR trace made at the time of installation is useful to determine where changes have occurred. Use of an OTDR for fault finding may require an experienced operator able to correctly judge the appropriate instrument settings to locate a problem accurately. This is particularly so in cases involving long distance, closely spaced splices or connectors, or PONs.

OTDRs are available with a variety of fiber types and wavelengths, to match common applications. In general, OTDR testing at longer wavelengths, such as 1550 nm or 1625 nm, can be used to identify fiber attenuation caused by fiber problems, as opposed to the more common splice or connector losses.

The dynamic range of an OTDR is limited by a combination of optical pulse output power, optical pulse width, input sensitivity, and signal integration time. Higher optical pulse output power, and better input sensitivity, combine directly to improve measuring range, and are usually fixed features of a particular instrument. However optical pulse width and signal integration time are user adjustable, and require trade-offs which make them application specific.

A longer laser pulse improves dynamic range and attenuation measurement resolution at the expense of distance resolution. For example, using a long pulse length, it may be possible to measure attenuation over a distance of more than 100 km, however, in this case, an optical event may appear to be over 1 km long. This scenario is useful for overall characterisation of a link, but would be of much less use when trying to locate faults. A short pulse length will improve distance resolution of optical events, but will also reduce measuring range and attenuation measurement resolution. The "apparent measurement length" of an optical event is referred to as the "dead zone". The theoretical interaction of pulse width and dead zone can be summarised as follows:

Pulse length Event dead zone
1 ns 0.15 m (theoretically)
10 ns 1.5 m (theoretically)
100 ns 15 m
1 µs 150 m
10 µs 1.5 km
100 µs 15 km
A stand for OTDR calibration (an optical fiber standard in the background)

The OTDR "dead zone" is a topic of much interest to users. Dead zone is classified in two ways. Firstly, an "Event Dead Zone" is related to a reflective discrete optical event. In this situation, the measured dead zone will depend on a combination of the pulse length (see table), and the size of the reflection. Secondly, an "Attenuation Dead Zone" is related to a non-reflective event. In this situation, the measured dead zone will depend on a combination of the pulse length (see table).

A long signal integration time effectively increases OTDR sensitivity by averaging the receiver output. The sensitivity increases with the square root of the integration time. For example, if the integration time is increased by 16 times, the sensitivity increases by a factor of 4. This imposes a practical sensitivity limit, with integration times of seconds to a few minutes.

The dynamic range of an OTDR is usually specified as the attenuation level where the measured signal becomes lost in the detection noise, for a particular combination of pulse length and signal integration time. This number is easy to deduce by inspection of the output trace, and is useful for comparison, but not in practice, since at this point, the measured values are random. So the practical measuring range is smaller, depending on required attenuation measurement resolution.

When an OTDR is used to measure the attenuation of multiple joined fiber lengths, the output trace can incorrectly show a joint as having gain, instead of loss. The reason for this is that adjacent fibers may have different backscatter coefficients, so the second fiber, with the same level of light travelling through it, reflects more light than the first fiber. If the OTDR is placed at the other end of this same fiber pair, it will measure an abnormally high loss at that joint. However if the two signals are then combined, the correct loss will be obtained. For this reason, it is common OTDR practice to measure and combine the loss from both ends of a link, so that cable joint, and end to end loss, can be measured accurately.

The theoretical distance measuring accuracy of an OTDR is extremely good, since it is based on software and a crystal clock with an inherent accuracy of better than 0.01%. This aspect does not need subsequent calibration since practical cable length measuring accuracy is typically limited to about 1% due to: cable length is not the same as the fiber length, the speed of light in the fiber is known with limited accuracy (the refractive index is only specified to 3 significant figures such as 1.45 etc.), and cable length markers have limited accuracy (0.5% – 1%).

An OTDR excels at identifying the existence of unacceptable point loss or return loss. Its ability to accurately measure absolute end-to-end loss or return loss can be quite poor, so cable acceptance usually includes an end-to-end test with a light source and power meter, and optical return loss meter. Its ability to exactly locate a hidden cable fault is also limited, so for fault-locating, it may be augmented with other localised tools such as a red laser fault locator, clip-on identifier, or "Cold Clamp" optical cable marker.

Reliability and quality of OTDR equipment[edit]

The reliability and quality of an OTDR should be based on its accuracy, measurement range, ability to resolve and measure closely spaced events, measurement speed, and ability to perform satisfactorily under various environmental extremes and after various types of physical abuse. In addition to cost, the instrument should be rated on features provided, size, weight, and ease of use.

Accuracy is defined as the correctness of the measurement (i.e., the difference between the measured value and the true value of the event being measured).

The measurement range of the OTDR is defined as the maximum attenuation that can be placed between the instrument and the event being measured, for which the instrument will still be able to measure the event within acceptable accuracy limits.

Instrument resolution is a measure of how close two events can be spaced and still be recognized as two separate events. The duration of the measurement pulse and the data sampling interval create a resolution limitation for OTDRs. The shorter the pulse duration and the shorter the data sampling interval, the better the instrument resolution, but the shorter the measurement range. Resolution is also often limited when powerful reflections return to the OTDR and temporarily overload the detector. When this occurs, some time is required before the instrument can resolve a second fiber event. Some OTDR manufacturers use a “masking” procedure to improve resolution. The procedure shields or “masks” the detector from high-power fiber reflections, preventing detector overload and eliminating the need for detector recovery.

Industry requirements for the reliability and quality of OTDRs are in GR-196, Generic Requirements for Optical Time Domain Reflectometer (OTDR) Type Equipment.

Types of OTDR-like test equipment[edit]

The common types of OTDR-like test equipment are:

  • Full-feature OTDR
  • Hand-held OTDR
  • Fiber Break Locator
  • RTU in RFTSs

The equipment is summarized below, and detailed in GR-196, Generic Requirements for Optical Time Domain Reflectometer (OTDR) Type Equipment.

  • Full-feature OTDR
    Full-feature OTDRs are traditional, optical time domain reflectometers. They are feature-rich and usually larger, heavier, and less portable than either the hand-held OTDR or the fiber break locator. Despite being characterized as large, their size and weight is only a fraction of that of early generation OTDRs. Often a full-feature OTDR has a main frame that can be fitted with multi-function plug-in units to perform many fiber measurement tasks. Larger color displays are common. The full-feature OTDR often has a greater measurement range than the other types of OTDR-like equipment. Often it is used in laboratories and in the field for difficult fiber measurements. Most full-feature OTDRs are powered from AC and/or a battery.
  • Hand-held OTDR and Fiber break locator
    Hand-held (formerly mini) OTDRs and fiber break locators are designed to troubleshoot fiber networks in a field environment, often using battery power. The two types of instruments cover the spectrum of approaches to fiber optic plant taken by communication providers. Hand-held, inexpensive (compared to full-feature) OTDRs are intended to be easy-to-use, light-weight, sophisticated OTDRs that collect field data and perform rudimentary data analysis. They may be less feature rich than full-feature OTDRs. Often they can be used in conjunction with PC-based software to perform data collection and sophisticated data analysis. Hand-held OTDRs are commonly used to measure fiber links and locate fiber breaks, points of high loss, high reflectance, end-to-end loss, and Optical Return Loss (ORL).
    Fiber break locators are intended to be low-cost instruments specifically designed to determine the location of a catastrophic fiber event, e.g., fiber break, point of high reflectance, or high loss. The fiber break locator is an opto-electronic tape measure designed to measure only distance to catastrophic fiber events.
    In general, hand-held OTDRs and fiber break locators are lighter and smaller, simpler to operate, and more likely to employ battery power than full-feature OTDRs. The intent with hand-held OTDRs and fiber break locators is to be inexpensive enough for field technicians to be equipped with one as part of a standard tool kit.
  • Remote Test Unit (RTU)
    The RTU is the testing module of the RFTS described in GR-1295, Generic Requirements for Remote Fiber Testing Systems (RFTSS). An RFTS enables fiber to be automatically tested from a central location. A central computer is used to control the operation of OTDR-like test components located at key points in the fiber network. The test components scan the fiber to locate problems. If a problem is found, its location is noted and the appropriate personnel are notified to begin the repair process. The RFTS can also provide direct access to a database that contains historical information of the OTDR fiber traces and any other fiber records for the physical fiber plant.
    Since OTDRs and OTDR-like equipment have many uses in the communications industry, operating environments vary widely, both indoors and outdoors. Most often, however, these test sets are operated in controlled environments, accessing the fibers at their termination points on fiber distribution frames. Indoor environments include controlled areas such as central offices (COs), equipment huts, or Controlled Environment Vaults (CEVs). Use in outside environments is rarer, but may include use in a manhole, aerial platform, open trench, or splicing vehicle.

OTDR Data Format[edit]

In the late 1990s, OTDR industry representatives and the OTDR user community developed a unique data format to store and analyze OTDR fiber data. This data was based on the specifications in GR-196, Generic Requirements for Optical Time Domain Reflectometer (OTDR) Type Equipment, referenced above. The goal was for the data format to be truly universal, in that it was intended to be implemented by all OTDR manufacturers. OTDR suppliers developed the software to implement the data format. As they proceeded, they identified inconsistencies in the format, along with areas of misunderstanding among users.

From 1997 to 2000, a group of OTDR supplier software specialists attempted to resolve problems and inconsistencies in what was then called the “Bellcore” OTDR Data Format. This group, called the OTDR Data Format Users Group (ODFUG), made progress. Since then, many OTDR developers continued to work with other developers to solve individual interaction problems and enable cross use between manufacturers.

In 2011, Telcordia decided to compile industry comments on this data format into one document entitled, SR-4731, Optical Time Domain Reflectometer (OTDR) Data Format. This Special Report (SR) summarizes the state of the Bellcore OTDR Data Format, renaming it as the Telcordia OTDR Data Format.

The data format is intended for all OTDR-related equipment designed to save trace data and analysis information. Initial implementations require standalone software to be provided by the OTDR supplier to convert existing OTDR trace files to the SR-4731 data format and to convert files from this universal format to a format that is usable by their older OTDRs. This file conversion software can be developed by the hardware supplier, the end user, or a third party. This software also provides backward compatibility of the OTDR data format with existing equipment.

The SR-4731 format describes binary data. While text information is contained in several fields, most numbers are represented as either 16-bit (2-byte) or 32-bit (4-byte) signed or unsigned integers stored as binary images. Byte ordering in this file format is explicitly low-byte ordering, as is common on Intel® processor-based machines. String fields are terminated with a zero byte “\0”. OTDR waveform data are represented as short, unsigned integer data uniformly spaced in time, in units of decibels (dB) times 1000, referenced to the maximum power level. The maximum power level is set to zero, and all waveform data points are assumed to be zero or negative (the sign bit is implied), so that the minimum power level in this format is -65.535 dB, and the minimum resolution between power level steps is 0.001 dB. In some cases, this will not provide sufficient power range to represent all waveform points. For this reason, the use of a scale factor has been introduced to expand the data point power range.[1]

See also[edit]

External links[edit]

 This article incorporates public domain material from the General Services Administration document "Federal Standard 1037C".

References[edit]

  1. ^ Terry Donovan, Senior Systems Engineer, Telcordia Technologies. Member of Optical Society of America, IEEE, "Optical Time Domain Reflectometer (OTDR) Data Format, SR-4731, Issue 2", July 2011