Free induction decay

From Wikipedia, the free encyclopedia

This is an old revision of this page, as edited by Koitus~nlwiki (talk | contribs) at 20:21, 8 October 2021. The present address (URL) is a permanent link to this revision, which may differ significantly from the current revision.

Free Induction Decay (FID) nuclear magnetic resonance signal seen from a well shimmed sample.

In Fourier transform nuclear magnetic resonance spectroscopy, free induction decay (FID) is the observable NMR signal generated by non-equilibrium nuclear spin magnetization precessing about the magnetic field (conventionally along z). This non-equilibrium magnetization can be created generally by applying a pulse of radio-frequency close to the Larmor frequency of the nuclear spins.

If the magnetization vector has a non-zero component in the xy plane, then the precessing magnetisation will induce a corresponding oscillating voltage in a detection coil surrounding the sample.[1] This time-domain signal (a sinusoid) is typically digitised and then Fourier transformed in order to obtain a frequency spectrum of the NMR signal i.e. the NMR spectrum.[2]

The duration of the NMR signal is ultimately limited by T2 relaxation, but mutual interference of the different NMR frequencies present also causes the signal to be damped more quickly. When NMR frequencies are well-resolved, as is typically the case in the NMR of samples in solution, the overall decay of the FID is relaxation-limited and the FID is approximately exponential (with the time constant T2 changed, indicated by T2*).[citation needed] FID durations will then be of the order of seconds for nuclei such as 1H.

Particularly if a limited number of frequency components are present, the FID may be analysed directly for quantitative determinations of physical properties, such as hydrogen content in aviation fuel, solid and liquid ratio in dairy products (time-domain NMR).[3]

Advances in the development of quantum-scale sensors, particularly NV centres, have enabled the observation of the FID of single nuclei.[4] When measuring the precession of a single nucleus, quantum mechanical measurement back action has to be considered. In this special case, also the measurement itself contributes to the decay as predicted by quantum mechanics.

References

  1. ^ Joseph P. Hornak. "The Basics of MRI". Rochester Institute of Technology. Chapter 4: NMR SPECTROSCOPY.
  2. ^ Duer, Melinda J. Introduction to Solid-State NMR Spectroscopy. Blackwell Publishing, 2004, p. 43-58.
  3. ^ H. Todt, G. Guthausen, W. Burk, D. Schmalbein, and A. Kamlowski. Water/moisture and fat analysis by time-domain NMR. Food Chemistry 96, 3 p. 436-440 (2006) doi: 10.1016/j.foodchem.2005.04.032
  4. ^ K. S. Cujia, J. M. Boss, K. Herb, J. Zopes, and C. L. Degen. Tracking the precession of single nuclear spins by weak measurements. Nature 571, 230-233 (2019) doi:10.1038/s41586-019-1334-9