Jump to content

Interference reflection microscopy

From Wikipedia, the free encyclopedia
Principle of interference reflection microscopy (IRM)
Principle of interference reflection microscopy (IRM) showing interference effect on reflected waves and the result on the final image intensity. Dark purple wave represents the light from the light source. The light purple waves are the reflections from the cell membrane and from the glass surface. Upon hitting the glass surface, the reflected waves are shifted half a wavelength. When the membrane is very close to the glass, the reflected light will be reflected out of phase with the reflected beam from the glass. This will cause destructive interference (see red line), resulting in a dark pixel. If there is more distance between the membrane and the glass, the returning waves will be less shifted and will cause constructive interference (see red line), resulting in a brighter pixel in the final image. Indicated are the typical refractive indices of the glass, medium and the cell membrane, which determine the amount of reflection.

Interference reflection microscopy (IRM), also called Reflection Interference Contrast Microscopy (RICM) or Reflection Contrast Microscopy (RCM) depending on the specific optical elements used, is an optical microscopy technique that leverages thin-film interference effects to form an image of an object on a glass surface. The intensity of the signal is a measure of proximity of the object to the glass surface. This technique can be used to study events at the cell membrane without the use of a (fluorescent) label as is the case for TIRF microscopy.

History and name

[edit]

In 1964, Adam S. G. Curtis coined the term Interference Reflection Microscopy (IRM), using it in the field of cell biology to study embryonic chick heart fibroblasts.[1][2] He used IRM to look at adhesion sites and distances of fibroblasts, noting that contact with the glass was mostly limited to the cell periphery and the pseudopodia.[1]

In 1975, Johan Sebastiaan Ploem introduced an improvement to IRM (published in a book chapter[3]), which he called Reflection Contrast Microscopy (RCM).[4] The improvement is to use a so-called anti-flex objective and crossed polarizers to further reduce stray light in the optical system. Today, this scheme is mainly referred to as Reflection Interference Contrast Microscopy (RICM),[5][6] the name of which was introduced by Bareiter-Hahn and Konrad Beck in 1979.[7]

However, the term IRM is sometimes used to describe an RICM setup. The multiplicity of names used to describe the technique has caused some confusion, and was discussed as early as 1985 by Verschueren.[8]

Theory

[edit]

To form an image of the attached cell, light of a specific wavelength is passed through a polarizer. This linear polarized light is reflected by a beam splitter towards the objective, which focuses the light on the specimen. The glass surface is reflective to a certain degree and will reflect the polarized light. Light that is not reflected by the glass will travel into the cell and be reflected by the cell membrane. Three situations can occur. First, when the membrane is close to the glass, the reflected light from the glass is shifted half of a wavelength, so that light reflected from the membrane will have a phase shift compared to the reflected light from the glass phases and therefore cancel each other out (interference). This interference results in a dark pixel in the final image (the left case in the figure). Second, when the membrane is not attached to the glass, the reflection from the membrane has a smaller phase shift compared to the reflected light from the glass, and therefore they will not cancel each other out, resulting in a bright pixel in the image (the right case in the figure). Third, when there is no specimen, only the reflected light from the glass is detected and will appear as bright pixels in the final image.

The reflected light will travel back to the beam splitter and pass through a second polarizer, which eliminates scattered light, before reaching the detector (usually a CCD camera) in order to form the final picture. The polarizers can increase the efficiency by reducing scattered light; however in a modern setup with a sensitive digital camera, they are not required.[9]

Theory

[edit]

Reflection is caused by a change in the refraction index, so on every boundary a part of the light will be reflected. The amount of reflection is given by the reflection coefficient , according to the following rule:[8]

Reflectivity is a ratio of the reflected light intensity () and the incoming light intensity ():[8]

Using typical refractive indices for glass (1.50–1.54, see list), water (1.31, see list), the cell membrane (1.48)[10] and the cytosol (1.35),[10] one can calculate the fraction of light being reflected by each interface. The amount of reflection increases as the difference between refractive indices increases, resulting in a large reflection from the interface between the glass surface and the culture medium (about equal to water: 1.31–1.33). This means that without a cell the image will be bright, whereas when the cell is attached, the difference between medium and the membrane causes a large reflection that is slightly shifted in phase, causing interference with the light reflected by the glass. Because the amplitude of the light reflected from the medium-membrane interface is decreased due to scattering, the attached area will appear darker but not completely black. Because the cone of light focused on the sample gives rise to different angles of incident light, there is a broad range of interference patterns. When the patterns differ by less than 1 wavelength (the zero-order fringe), the patterns converge, resulting in increased intensity. This can be obtained by using an objective with a numerical aperture greater than 1.[8]

Requirements

[edit]

In order to image cells using IRM, a microscope needs at least the following elements: 1) a light source, such as a halogen lamp, 2) an optical filter (which passes a small range of wavelengths), and 3) a beam splitter (which reflects 50% and transmits 50% of the chosen wavelength).

The light source needs to produce high intensity light, as a lot of light will be lost by the beam splitter and the sample itself. Different wavelengths result in different IRM images; Bereiter-Hahn and colleagues showed that for their PtK 2 cells, light with a wavelength of 546 nm resulted in better contrast than blue light with a wavelength of 436 nm.[7] There have been many refinements to the basic theory of IRM, most of which increase the efficiency and yield of the image formation. By placing polarizers and a quarter wave plate between the beam splitter and the specimen, the linear polarized light can be converted into circular polarized light and afterwards be converted back to linear polarized light, which increases the efficiency of the system. The circular polarizer article discusses this process in detail. Furthermore, by including a second polarizer, which is rotated 90° compared to the first polarizer, stray light can be prevented from reaching the detector, increasing the signal to noise ratio (see Figure 2 of Verschueren[8]).

Biological applications

[edit]

There are several ways IRM can be used to study biological samples. Early examples of uses of the technique focused on cell adhesion[1] and cell migration.[11]

Vesicle fusion

[edit]
Two chromaffin cells imaged with DIC (left) and IRM (right).
Vesicle fusion visualized using IRM. Scale bar represents 2 μm.

More recently, the technique has been used to study exocytosis in chromaffin cells.[9] When imaged using DIC, chromaffin cells appear as round cells with small protrusions. When the same cell is imaged using IRM, the footprint of the cell on the glass can be clearly seen as a dark area with small protrusions. When vesicles fuse with the membrane, they appear as small light circles within the dark footprint (bright spots in the top cell in the right panel).

An example of vesicle fusion in chromaffin cells using IRM is shown in movie 1. Upon stimulation with 60 mM potassium, multiple bright spots begin to appear inside the dark footprint of the chromaffin cell as a result of exocytosis of dense core granules. Because IRM doesn't require a fluorescent label, it can be combined with other imaging techniques, such as epifluorescence and TIRF microscopy to study protein dynamics together with vesicle exocytosis and endocytosis. Another benefit of the lack of fluorescent labels is reduced phototoxicity.

References

[edit]
  1. ^ a b c Curtis AS (February 1964). "THE MECHANISM OF ADHESION OF CELLS TO GLASS : A Study by Interference Reflection Microscopy". The Journal of Cell Biology. 20 (2): 199–215. doi:10.1083/jcb.20.2.199. PMC 2106393. PMID 14126869.
  2. ^ Filler TJ, Peuker ET (April 2000). "Reflection contrast microscopy (RCM): a forgotten technique?". The Journal of Pathology. 190 (5): 635–8. doi:10.1002/(SICI)1096-9896(200004)190:5<635::AID-PATH571>3.0.CO;2-E. PMID 10727991. S2CID 43800311.
  3. ^ Furth, Ralph van (1975). "Chapter 27: Reflection-contrast microscopy as a tool for investigation of the attachment of living cells to a glass surface". Mononuclear Phagocytes: In Immunity, Infection, and Pathology. Oxford: Blackwell Scientific. pp. 405–421. ISBN 0-632-00471-1. OCLC 2701405.
  4. ^ Ploem, Johan Sebastiaan (2019-02-21). "Applications of reflection‐contrast microscopy, including the sensitive detection of the results ofin situhybridisation a review". Journal of Microscopy. 274 (2). Wiley: 79–86. doi:10.1111/jmi.12785. ISSN 0022-2720. PMID 30720204. S2CID 73431526.
  5. ^ Theodoly, O.; Huang, Z.-H.; Valignat, M.-P. (2009-11-30). "New Modeling of Reflection Interference Contrast Microscopy Including Polarization and Numerical Aperture Effects: Application to Nanometric Distance Measurements and Object Profile Reconstruction" (PDF). Langmuir. 26 (3). American Chemical Society (ACS): 1940–1948. doi:10.1021/la902504y. ISSN 0743-7463. PMID 19947618.
  6. ^ Limozin, Laurent; Sengupta, Kheya (2009-11-03). "Quantitative Reflection Interference Contrast Microscopy (RICM) in Soft Matter and Cell Adhesion". ChemPhysChem. 10 (16). Wiley: 2752–2768. doi:10.1002/cphc.200900601. ISSN 1439-4235.
  7. ^ a b Bereiter-Hahn J, Fox CH, Thorell B (September 1979). "Quantitative reflection contrast microscopy of living cells". The Journal of Cell Biology. 82 (3): 767–79. doi:10.1083/jcb.82.3.767. PMC 2110483. PMID 389938.
  8. ^ a b c d e Verschueren H (April 1985). "Interference reflection microscopy in cell biology: methodology and applications". Journal of Cell Science. 75 (1): 279–301. doi:10.1242/jcs.75.1.279. PMID 3900106.
  9. ^ a b Wu MM, Llobet A, Lagnado L (November 2009). "Loose coupling between calcium channels and sites of exocytosis in chromaffin cells". The Journal of Physiology. 587 (Pt 22): 5377–91. doi:10.1113/jphysiol.2009.176065. PMC 2793871. PMID 19752110.
  10. ^ a b Dunn, Andrew Kenneth (1997). "Cell Structure". Light scattering properties of cells (PhD thesis). University of Texas at Austin. OCLC 39949488. Retrieved February 23, 2010.
  11. ^ Godwin SL, Fletcher M, Burchard RP (September 1989). "Interference reflection microscopic study of sites of association between gliding bacteria and glass substrata". Journal of Bacteriology. 171 (9): 4589–94. doi:10.1128/jb.171.9.4589-4594.1989. PMC 210255. PMID 2768185.

Further reading

[edit]
  • Limozin, Laurent; Sengupta, Kheya (2009-11-09). "Quantitative Reflection Interference Contrast Microscopy (RICM) in Soft Matter and Cell Adhesion". ChemPhysChem. 10 (16): 2752–2768. doi:10.1002/cphc.200900601. ISSN 1439-4235.
[edit]