Jump to content

Clearing factor

From Wikipedia, the free encyclopedia

This is an old revision of this page, as edited by BD2412 (talk | contribs) at 14:29, 11 September 2010 (Fixing links to disambiguation pages, improving links, other minor cleanup tasks using AWB). The present address (URL) is a permanent link to this revision, which may differ significantly from the current revision.

In centrifugation the clearing factor or k factor represents the relative pelleting efficiency of a given centrifuge rotor at maximum rotation speed. It can be used to estimate the time (in hours) required for sedimentation of a fraction with a known sedimentation coefficient (in svedbergs):

The value of the clearing factor depends on the maximum angular velocity of a centrifuge (in rad/s) and the minimum and maximum radius of the rotor:

As the rotational speed of a centrifuge is usually specified in RPM, the following formula is often used for convenience:

Centrifuge manufacturers usually specify the minimum, maximum and average radius of a rotor, as well as the factor of a centrifuge-rotor combination.

For runs with a rotational speed lower than the maximum rotor-speed, the factor has to be adjusted:

2

References